Annually, Sheath filter – Luminex 200 User Manual with IS 2.3 Software User Manual

Page 37

Advertising
background image

x

MAP Technology

Maintenance and Cleaning

PN 89-00002-00-144 Rev. B

3 - 11

Annually

Sheath Filter

‹

To change the Luminex 200 analyzer sheath filter:

1. Disconnect the Luminex 200 analyzer from ac power by turning

off the power switch on the rear of the analyzer, then unplugging
the analyzer power cord from the wall source.

2. Disconnect the sheath fluid tubing before changing the filter.

3. Open the left door on the Luminex 200 analyzer. Disconnect the

filter by pushing down on the metal clamps on each connection.
See Figure 3-10.

Figure 3-10 Sheath Filter

4. Connect the new sheath filter, matching up the color-coded

fittings. The arrow on the sheath filter should be pointing up.

5. Reconnect the sheath fluid tubing.

6. Plug in and power on the Luminex 200 analyzer.

7. Close the left analyzer door.

8. Prime twice.