A normal bead detail display is shown below. It depicts a tight bead
population within a white region.

Figure 4-1 Normal Bead Detail

The histogram for the bead detail above looks like this:

Figure 4-2 Normal Histogram

Symptom

Possible problem

Solution

xMAP microspheres classify
too high
.

You may be using
photobleached calibration
microspheres.

Replace the calibration
microspheres with a fresh batch.
To avoid photobleaching,
protect your microspheres from