Luminex 200 System User Manual

x

MAP Technology

3 - 12

PN 89-00002-00-144 Rev. B

As required

Fuses

‹

To replace the fuses in either the Luminex 200 analyzer or the
Luminex XYP instrument:

1. Turn off the power switch on the rear of the analyzer or

instrument, then unplug the instrument power cord from the wall
source. Remove the power cord from the analyzer or instrument.

2. With a small, flathead screwdriver, open the module door on the

lower left corner of the back of the analyzer or instrument. See
Figure 3-11.

Figure 3-11 Opening the Module Door

3. Remove the red cartridge (use a flathead screwdriver).

4. Check both fuses for damage.

5. Replace damaged fuses with the type specified on the sticker to

the right of the power input module.

6. Replace the module door.

7. Plug in and power on the analyzer or instrument.

Danger:

To avoid serious injury or death by electric shock,

you must turn off the system and unplug it from the wall
outlet.