2 Important

Safety

Advices

2.1 Important Safety Advice

Warning: Make sure to read the following advice prior to installations of the Roboocyte2.
If you do not fulfill all requirements stated below, this may lead to malfunctions, breakage,
or even fatal injuries. Obey always the rules of local regulations and laws. Only qualified
personnel should be allowed to perform laboratory work. Work according to good laboratory
practice to obtain best results and to minimize risks.

The product has been built to the state of the art and in accordance with recognized safety
engineering rules. The device may only

 be used for its intended purpose;

 be used when in a perfect condition.

Improper use could lead to serious, even fatal injuries to the user or third parties and damage
to the device itself or other material damage.

Malfunctions which could impair safety should be rectified immediately.

2.1.1 High

Voltage

Electrical cords must be properly laid and installed. The length and quality of the cords must
be in accordance with local provisions.

Only qualified technicians may work on the electrical system. It is essential that the accident
prevention regulations and those of the employers' liability associations are observed.

Each time before starting up, make sure that the mains supply agrees with the specifications
of the product.

Check the power cord for damage each time the site is changed. Damaged power cords should
be replaced immediately and may never be reused.

Check the leads for damage. Damaged leads should be replaced immediately and may never
be reused.

Liquids may cause short circuits or other damage. Keep the power supply and the power cords
always dry. Do not handle it with wet hands.

2.1.2 Requirements for the Installation

The Roboocyte2 weighs more than 22 kg. Always grip it tightly and do not carry it alone,
but with the aid of another person.

The movement of the well plate carrier can lead to vibrations of the workbench on which the
Roboocyte2 is set up. Therefore, the Roboocyte2 must be set up on a rigid, vibration-free base.
The base must also be sufficiently solid to carry the weight of the device.

The Roboocyte2 should be operated only in an air conditioned room. A room temperature
of 20 °C (or less) is recommended. Make sure that the device is not subject to direct sunlight.
It may overheat.

If the air cannot circulate freely around the external power supply, the device may overheat.
Do not shield the power supply by laying anything on top of it.

The external power supply is only for use with the Roboocyte2. Do not connect it to any other