Roboocyte2 Manual

74

If you need to start the alignment again, for example if you have moved the carrier accidentally
by hand, click "Cancel" and restart the alignment process from the beginning.

Click "Next". The current position is saved, and the old alignment position is overwritten.
All following movements will be related to the new position now.

Warning: Do not click "Next" if the alignment has not been finished. This could lead to wrong
movements of the Roboocyte2, which are potentially hazardous. If you want to cancel the
alignment, leave this dialog box by clicking "Cancel" or "Back".

The Resistance Test dialog box opens. You can now test the electrodes resistance. Please read
the next chapter "Testing the Electrode Impedance".

7.3 Testing the Electrode Resistance

Before starting to do experiments with a new measuring head, the electrode resistance of both
glass microelectrodes should be checked. After finishing the alignment the "Resistance Test"
window of the "Alignment Wizard" appears..