Testing the Noise Level

The current noise level during voltage clamp can be evaluated by recording from the TMC at
different sampling rate and gain settings. In the following, holding currents at 0 mV holding
potential are displayed to give you an impression about the expected noise levels.

Gain 1000/100 - sampling rate 20 kHz

Gain 1000/100 - sampling rate 10 kHz

Gain 1000/100 - sampling rate 1 kHz