Multichannel Systems Roboocyte2 Manual User Manual
Page 34
Advertising
Roboocyte2 Manual
30
Testing the Noise Level
The current noise level during voltage clamp can be evaluated by recording from the TMC at
different sampling rate and gain settings. In the following, holding currents at 0 mV holding
potential are displayed to give you an impression about the expected noise levels.
Gain 1000/100 - sampling rate 20 kHz
Gain 1000/100 - sampling rate 10 kHz
Gain 1000/100 - sampling rate 1 kHz
Advertising