Setting Up Hardware and Software

29

Press the two autoscale arrows at the bottom of the data plot window

.

This will bring the current traces completely on the display.

You can also zoom into a certain region by using the zoom box tool

.

Complete zoom-out can be accomplished by pressing the zoom-out button

.

The current traces displayed now have been recorded with a proportional gain of 1000 nA/mV
and an integral gain of 100 1/s at a sampling rate of 20 kHz. With the test model cell, this leads
to voltage-clamp within less than 1 ms. Similar values can be expected with a "tight" non-injected
oocyte.