6. Click the alignment icon
in the main window. The "Preparation" window
of the "Alignment Wizard" dialog appears.
7. Click the button “Change Probe Pos”, The z-axis will move down a bit now allowing
better access to the place where the measuring head electrode wires will be connected to.
8. Fix the measuring head on the cone of the z-axis.
9. Use the provided forceps to plug the four connectors into the appropriate sockets on the axis.
Connectors, viewed from the front and from left to right: Current electrode (left glass electrode),
bath reference for the current electrode (RI), voltage electrode (right glass electrode), bath
reference for the voltage electrode (RU). Attach the perfusion manifold directly to the inlet
nozzle.
10. Attach the tubings coming from either Roboflow or from the Gilson liquid handler to the
measuring head. The suction tube outlet is the coming out of the measuring head below
the inlet tube.