8-3

8. EXPLOSION PROTECTED TYPE INSTRUMENT

(Remote Flowtube)

WARNING

• De-energize before opening.
• Take care not to generate mechanical spark

when access to the instrument and peripheral
devices in hazardous locations.

(5) Maintenance and Repair

WARNING

The instrument modification or parts replacement
by other than authorized representative of
Yokogawa Electric Corporation is prohibited and
will void the certification.

(6) Data Plate

(Integral Flowmeter)

VDC 0.2A MAX.

mA (0-750

)

VDC 12W

VAC 50/60Hz 30VA 12W

0344

0038

2G
1D

Made in

*3)

MPa MAX.

mm

˚C

˚C

L

H

WARNING

User’s Manual

MAX.PROCESS TEMP.(˚C)

+130

T3

+120

T4

+85

T5

+70

T6

T110

°C

T100

°C

T85

°C

T75

°C

TEMP. CLASS

No.: KEMA03ATEX2435

EEx dme [ia] C T6...T3

ENCLOSURE: IP66, IP67

ELECTRODE CIRCUIT Um: 250Vac/dc

*2)
*4)

MAX. SURFACE TEMP.

FOR DUST-PROOF

Tamb: –40 TO +60

°C

TOKYO 180-8750 JAPAN

(Remote Flowtube)

MPa MAX.

˚C

˚C

L

H

COMB.NO.

mm

TOKYO 180-8750 JAPAN

2G
1D

F9805AG

WARNING

User’s Manual

MAX.PROCESS TEMP.(˚C)

+150

T3

+120

T4

+85

T5

+70

T6

T115

°C

T100

°C

T85

°C

T75

°C

TEMP. CLASS

No.: KEMA03ATEX2435
EEx dme [ia] C T6...T3
ENCLOSURE: IP66,IP67
ELECTRODE CIRCUIT Um: 250Vac/dc
Tamb: –40 TO +60

°C

Made in

*3)

MAX. SURFACE TEMP.

FOR DUST-PROOF

0344
0038

*2)
*4)

MODEL: Specified model code
SUFFIX: Suffix codes of the model code
STYLE: Specified style code
SIZE: Nominal size of apparatus
METER FACTOR: Sensor constant number of apparatus
SUPPLY: Power supply voltage of apparatus
OUTPUT: Output signal of apparatus
FLUID TEMP.: Fluid temperature of apparatus
FLUID PRESS: Fluid pressure of apparatus
AMB. TEMP., Tamb: Ambient temperature
No.: Manufacturing serial number *1)
CE: CE marking
II 2G: Group II Category 2 Gas atmosphere
II 1D: Group II Category 1 Dust atmosphere

No

.

: KEMA 03ATEX2435:

EC Type Examination certificate number

EEx dme[ia]IIC T6...T3: Protection type and temp. class
ELECTRODE CIRCUIT Um: Voltage of electrode circuit
ENCLOSURE: Enclosure protection code

WARNING

: Warning to apparatus

YOKOGAWA

᭜ TOKYO 180-8750 JAPAN :

Name and address of manufacturer

*1) The third figure from the last shows the last one

figure of the year of production. For example, the
year of production of the product engraved as
follows is year 2005.

No. S5EA05158 545

Produced in 2005

*2) The identification number of the notified body :

0344 KEMA Netherland

*3) The product-producing country
*4) In case of the sizes of 2.5 to 25mm (0.1 to 1.0

in.) , “0038” is not described.