IM 01E20D01-01E

4-12

4. WIRING

A
B
C

EX1
EX2

Terminal

Symbols

Description

F042401.EPS

Flow signal output

Excitation current input

Functional grounding
(Only for explosion proof type)

Protective grounding
(Oitside of the terminal)

Figure 4.2.10 Terminal Configuration (Explosion proof

Type, Size 2.5 to 400 mm (0.1 to 16 in.))

F040211.EPS

Terminal

Symbols

A
B
C

EX1
EX2

Flow signal output

Excitation current input

Protective grounding
(Outside of the terminal)

Description

Figure 4.2.11 Terminal Configuration (General-purpose

Use, Submersible Type, Size 500 to 1000
mm (20 to 40 in.))

F040212.EPS

Terminal

Symbols

A
B
C

EX1
EX2

Flow signal output

Excitation current input

Protective grounding
(Outside of the terminal)

Description

C

B

A

EX2

EX1

Figure 4.2.12 Terminal Configuration (General-purpose

Use, submersible Type, Size 1100 to 2600
mm (44 to 104 in.))

(3) Wiring the Remote Flowtube (General-

Purpose Use, Submersible Type, Sanitary
Type, Size 2.5 to 400 mm (0.1 to 16 in.))
with Converters

1) Connection with the AXFA11 converter

Connect wiring as shown in the figure below.

I+

I–

CURRENT OUT

AL+

AL–

C

SA

A

B

SB

ALARM OUT

N/–

L/+

POWER SUPPLY

EX2

EX1

EXCIT ATION

P–

SI1+

SI2+

COM

P+

PULSE OUT

STATUS IN

SIGNAL

SO1+

COM

SO2+

STATUS OUT

FUSE

2.5A 250V

AXFA11 converter

Remote flowtube

AXFC dedicated
signal cable

Converter

Remote

flowtube

F0426.EPS

SA

A

B

SB

C

EX1

EX2

Taping*

A

B

Taping*

C

EX1

EX2

* Individually tape and insulate the

shields corresponding to SA and
SB on the remote flowtube side.

EX2

EX1

A

B

C

Excitation cable

Figure 4.2.13 Wiring Diagram

2) Connection with the AXFA14 converter

Connect wiring as shown in the figure below.

AXFC dedicated
signal cable

EX2

EX1

A

B

C

Excitation cable

EX1

EX2

C

SA

A

B

SB

F0427.EPS

Converter

Remote

flowtube

SA

A

B

SB

C

EX1

EX2

Taping*

A

B

Taping*

C

EX1

EX2

* Individually tape and insulate the

shields corresponding to SA and
SB on the remote flowtube side.

AXFA14 converter

Remote flowtube

Figure 4.2.14 Wiring Diagram