Maintenance, 1 changing direction ofelectrical connection, Maintenance -1 – Yokogawa ADMAG AXF User Manual

Page 50: Changing direction of electrical connection -1, 1 changing direction of electrical connection

Advertising
background image

IM 01E20D01-01E

5-1

5. MAINTENANCE

5.

MAINTENANCE

WARNING

• Maintenance work must be carried out by the

trained personnel having knowledge of safety
standard and not by operators.

• When opening the cover, wait for more than 10

minutes after turning off the power. Further-
more, opening of the cover must also be
carried out by the trained personnel having
knowledge of safety standard.

CAUTION

• Explosion protected type must be, as a rule,

removed to a non-hazardous area for mainte-
nance and be disassembled and reassembled
to the original state.

• The cover is locked by the special screw. In

case of opening the cover, use the hexagonal
wrench attached.

• Be sure to lock the cover by the special screw

using the hexagonal wrench attached after
installing the cover.

5.1 Changing Direction of

Electrical Connection

IMPORTANT

The following types can not be changed direction
of electrical connection after delivery.
• Submersible Type.
• Optional code DHC (for district heating and

cooling or condensation-proof).

• Size of 1100 mm (44 in.) or larger.

(1) The following tools are required to change the

direction of the electrical connection:

• Hexagonal wrench (nominal size 1.5): Comes

with the instrument.

• Wrench

(2) Turn off the power to the flowmeter.

(3) Using the wrench, loosen the hexagonal nut at the

neck of the instrument.

F0501.EPS

(4) Using the hexagonal wrench, loosen the screw in

the neck.

(5) Turn the converter or the terminal box in the

desired direction.

NOTE

The converter and the terminal box can be
turned –140 degree to +180 degree from the
arrow mark indicating the flow direction. Do not
exceed these angle.

(6) Using the hexagonal wrench, retighten the neck

screw.

F0502.EPS

(7) Using the wrench, retighten the hexagonal nut at

the neck. After that, check that the converter or
terminal box is fixed.

Advertising