Accessories, Terminal configuration and terminal wiring, Integral flowmeter – Yokogawa ADMAG AXF User Manual

Page 78: Remote flowtube

Advertising
background image

IM 01E20D01-01E

6-16

6. OUTLINE

ACCESSORIES

Remote Flowtube (size 2.5 to 1000 mm(0.1 to 40 in.)):

Centering device (wafer type only): 1 pc.
Hexagonal wrench: 2 pcs.

Integral Flowmeter:

Centering device (wafer type only): 1 pc.
Hexagonal wrench: 2 pcs.

TERMINAL CONFIGURATION AND

TERMINAL WIRING

Integral Flowmeter “

+

SUPPLY

DIO

POWER

-

+

/

LN

/

I

-

+

+

-

DO

-

Terminal

Symbols

Functional grounding

Power supply

Current output 4 to 20mA DC

Pulse output/Alarm output/

Status output

Alarm output/Status output

Status input

Protective grounding

(Outside of the terminal)

Description

N/-
L/+
I+
I-
DO+
DO-
DIO+
DIO-

Terminal configuration

Terminal wiring

F41.EPS

Remote Flowtube

F42.EPS

Terminal wiring

Terminal

Symbols

A
B
C

EX1
EX2

Flow signal output

Excitation current input

Functional grounding
(Only for explosion proof type)

Description

Protective grounding
(Outside of the terminal)

(Only for Explosion proof type)

size 2.5 to 400 mm (0.1 to 16 in.)

size 500 to 1000 mm (20 to 40 in.)

size 1100 to 2600 mm (44 to 104 in.)

EX2

EX1

C

B

A

EX2

EX1

C

B

A

Terminal configuration

C

B

A

EX2

EX1

EX2

EX1

C

B

A

Note: When submersible type or option code DHC is

selected, waterproof glands and a 30m long cable are
attached.