Normal operating conditions – Yokogawa ADMAG AXF User Manual

Page 73

Advertising
background image

IM 01E20D01-01E

6-11

6. OUTLINE

CAUTION

*1: When performing the Insulation Resistance Test or the

Withstand Voltage Test, please obey the following
caution.

• Following the relevant test, wait for more than 10

seconds after the power supply has been turned off
before removing the cover.

• Remove all wires from terminals before testing.
• When the power terminal has a lighting protector

(optional code A), remove the short bar at the ground
terminal.

• After testing, be sure to discharge by using a resistance

and return all wires and the short bar to its correct
position.

• Screws must be tightened to a torque of 1.18 N-m or

more.

• After closing the cover, the power supply can be

restored.

Safety Requirement Standards:

EN61010-1

• Altitude at installation site: Max. 2000 m above sea leve
• Installation category based on IEC1010:

Overvoltage category II (“II” applies to electrical equip-
ment which is supplied from the fixed installation like
distribution board.)

• Pollution degree based on IEC1010

Pollution degree 2 (“Pollution degree” describes the
degree to which a solid, liquid, or gas which deteriorates
dielectric strength or surface resistivity is adhering. “2”
applies to a normal indoor atmosphere.)

EMC Conformity Standards:

,

EN61326-1 Class A, Table 2 (For use in industrial locations)
EN61326-2-3
EN61000-3-2 ClassA
EN61000-3-3

CAUTION

This instrument is a class A product, and it is designed for
use in the industrial environment.
Please use this instrument in the industrial environment
only.

NORMAL OPERATING CONDITIONS

Ambient Temperature: –40

° to +60°C (–40° to +140°F)

*1: Minimum temperature should also be limited according

to minimum fluid temperature of flow tube’s
specification.
Refer to description of “Fluid Temperature and
Pressure”.

*2: Indicator’s operating range (integral flowmeter): –20

° to

+60

°C (–4° to +140°F)

*3: Maximum temperature should be +50

°C (+122°F) in the

case of power supply code 2 (integral flowmeter).

Ambient Humidity: 0 to 100%

Lengthy continuous operation at 95% or more is not
recommended.

Power Supply (integral flowmeter):

Power supply code 1:

• AC specifications

Rated power supply: 100 to 240 V AC, 50/60 Hz
(Operating voltage range: 80 to 264 V AC)

• DC specifications

Rated power supply: 100 to 120 V DC
(Operating voltage range: 90 to 130 V DC)

Power supply code 2:

• AC specifications

Rated power supply: 24 V AC, 50/60 Hz
(Operating voltage range: 20.4 to 28.8 V AC)

• DC specifications

Rated power supply: 24 V DC
(Operating voltage range: 20.4 to 28.8 V DC)

Supply Voltage and

Cable Length for Power Supply Code 2

0

20.4

22

24

26

28.8

200 ( 660)

446 (1460)

600 (1970)

796 (2610)

F01.EPS

Cable cross section area: 1.25 mm

2

Cable cross section area: 2 mm

2

Allo

w

ab

le cab

le length m(ft)

Usable range E (V)

Fluid Conductivity:

Size 2.5 to 10 mm (0.1 to 0.4 in.): 5

µS/cm or larger

Size 15 to 125 mm (0.5 to 5 in.): 1

µS/cm or larger

Size 150 to 400 mm (6 to 16 in.): 3

µS/cm or larger

Note: For a fluid with large flow noise (pure water, pure

alcohol or others), or a fluid wth low conductivity and
low viscosity, the output fluctuates and is impossible
to measure accurately.

Size 500 to 2600 mm (20 to 104 in.) : 50

µS/cm or larger.