Yokogawa ADMAG AXF User Manual
Page 59
Advertising
IM 01E20D01-01E
5-10
5. MAINTENANCE
Coil Circuit
Checking is possible even if the pipe is filled with
fluid.
Test Terminals
Test Voltage
Specification
500 V DC (Use an insulation
tester or the equivalent.)
1 M
⍀ or more
Between terminals
EX1 and C
T050601.EPS
Signal Circuit
Before testing, be sure to empty and dry the interior of
the pipe, checking that there is no adhesive material.
Also undo the wiring connection on the converter side
before testing.
Test Terminals
Test Voltage
Specification
Between terminals
A and C
Between terminals
B and C
500 V DC (Use an insulation
tester or the equivalent.)
100 M
⍀ or
more for each
T050602.EPS
Advertising