IM 01E20D01-01E

5-10

5. MAINTENANCE

Coil Circuit

Checking is possible even if the pipe is filled with
fluid.

Test Terminals

Test Voltage

Specification

500 V DC (Use an insulation
tester or the equivalent.)

1 M

⍀ or more

Between terminals
EX1 and C

T050601.EPS

Signal Circuit

Before testing, be sure to empty and dry the interior of
the pipe, checking that there is no adhesive material.
Also undo the wiring connection on the converter side
before testing.

Test Terminals

Test Voltage

Specification

Between terminals
A and C
Between terminals
B and C

500 V DC (Use an insulation
tester or the equivalent.)

100 M

⍀ or

more for each

T050602.EPS