Yokogawa Button Operated MV2000 User Manual

Page 66

Advertising
background image

3-7

IM MV1000-17E

Using the Serial Interface

1

2

3

4

5

6

7

App

Index

Note

• The way to eliminate noise varies depending on the situation. In the connection example, the

cable shield is connected only to the MV’s ground (one-sided grounding). This is effective
when there is a difference in the electric potential between the PC’s ground and the MV’s
ground, which may be the case with long distance communications. If there is no difference
in the electric potential between the PC’s ground and the MV’s ground, connecting the cable
shield to ground at both ends may be effective (two-sided grounding). In some cases, using
two-sided grounding with a capacitor connected in series at one end is effective. Consider
these possibilities to eliminate noise.

• When using the two-wire system (Modbus protocol), the 485 driver must be set to high

impedance within 3.5 characters after the last data byte that the host PC sends.

Serial Interface Converter

We recommend the following converter.

MODEL RC-770X by SYSMEX RA CO.,LTD; SI-30FA by LINE EYE; or ML2 by
YOKOGAWA.

CAUTION

In converters other than those that we recommend, the FG and SG terminals may
not be isolated. In such case, do not follow the diagram on the previous page (do
not connect anything to the FG and SG terminals). Especially in long distance
communications, the potential difference that occurs may damage the instruments
or may cause communication errors. For converters that do not have the SG
terminal, they can be used without the signal ground. For details, see the converter
manual.

In converters other than those that we recommend, the signal polarities may be reversed

(A/B or +/- indication). In this case, reverse the connection.

In the case of a two-wire system, the host device must control the converter transmission

driver to prevent collisions of transmitted and received data. If you are using one of the

recommended converters, control the transmission driver using the RS (RTS) signal on

the RS-232.

When the System Contains Instruments That Only Support the RS-422

Interface

In a four-wire system, up to 32 MVs can connect to a single host device. However, this

may not be possible if the system contains instruments that support only the RS-422

interface.

When the System Contains YOKOGAWA Recorders That Only Support the

RS-422 Interface

Only up to 16 instruments can be connected. Some of the conventional YOKOGAWA

recorders (HR2400 and μR, for example) only support the RS-422 driver. If the system

contains these recorders, only up to 16 instruments can be connected.

Note

In the RS-422 standard, the maximum number of connections that are allowed on one port is
10 (for a four-wire system).

Terminator

In a multidrop connection (including point-to-point connection), connect a terminator to

the MV if the MV is connected to the end of the chain. Do not connect a terminator to a

MV in the middle of the chain. In addition, turn the host device terminator ON (see the

host device manual). If a converter is being used, turn ON its terminator. We recommend

converters with a built-in terminator.

Select the appropriate terminator (120 Ω), indicated in the figure, according to the

characteristic impedance of the line, the installation conditions of the instruments, and so

on.

3.2 Connecting the MV

Advertising
This manual is related to the following products: