Sp sets the partial expanded display, St sets a tag, Sx sets a group – Yokogawa Button Operated MV2000 User Manual

Page 88: Sl sets a trip line

Advertising
background image

4-14

4.4 Setting Commands (Setting)

IM MV1000-17E

Query

SZ[ p1]?

Example Display channel 002 in a 30%-to-50% zone.

SZ002,30,50

Description • You can specify a computation channel (or

make a query) on models with the /M1 math
option. You can specify an external input
channel (or make a query) on models with the
/MC1 external input channel option.

• The width of the waveform display area along

the amplitude axis is assumed be 100%.

• The zone width must be at least 5%.
• Set the top edge of the zone to a value greater

than the bottom edge of the zone.

SP

Sets the partial expanded display

Syntax

SP p1,p2,p3,p4<terminator>
p1 Measurement, computation, or external

input channel number

p2 Partial expanded setting ON/OFF
p3 Boundary position (1 to 99) [%]
p4 Boundary value

Query

SP[ p1]?

Example Partially expand the display of channel 001. Set

the boundary position to 25% and the boundary
value to 1.00 V.
SP001,ON,25,100

Description • You can specify a computation channel (or

make a query) on models with the /M1 math
option. You can specify an external input
channel (or make a query) on models with the
/MC1 external input channel option.

• Parameter p2 cannot be set to ON if the input

range (SR command) is set to SKIP.

• Parameter p2 cannot be set to ON for a

computation channel if computation is OFF
(SO command).

• Parameter p2 cannot be set to ON for an

external input channel if the channel is OFF
(ER command).

• The range between the span upper and

lower limits (scale upper and low limits when
scaling is enabled) is assumed to be 100% for
parameter p3.

• You can set p4 in the range of (span upper

limit – 1) to (span lower limit + 1). If scaling is
enabled, you can set p4 in the range of (scaling
upper limit – 1) to (scaling lower limit + 1).

• The decimal place and the number of digits

are set to the same values as the span and
scaling settings (see the SR command).

• You can execute this command (or make a

query) when the partial expanded display
function is set to USE (XU command).

• You cannot execute this command if the

partial expanded display range is not available
(such as when the span width is set to 1).

ST

Sets a tag

Syntax

ST p1,p2<terminator>
p1 Measurement, computation, or external

input channel number

p2 Tag (up to 16 characters)

Query

ST[ p1]?

Example Set the channel 002 tag to TAG2.

ST002,TAG2

Description • For the characters that can be used in a tag,

see appendix 1, “ASCII Character Codes.”
However, you cannot use a semicolon or
comma in a tag.

• You can specify a computation channel (or

make a query) on models with the /M1 math
option. You can specify an external input
channel (or make a query) on models with the
/MC1 external input channel option.

SX

Sets a group

Syntax

SX p1,p2,p3<terminator>
p1 Group number
p2 Group name (up to 16 characters)
p3 Channel configuration

Query

SX[ p1]?

Example Assign channels 001, 003, 004 to 006 to group

number 1 and assign GROUP2 for the group
name.
SX1,GROUP2,001.003.004-006
Assign channels by using a period to separate
each channel or a hyphen to specify a range of
channels.

Description For the characters that can be used in a group

name, see appendix 1, “ASCII Character Codes.”
However, you cannot use a semicolon or comma
in a group name.

SL

Sets a trip line

Syntax

SL p1,p2,p3,p4,p5,p6<terminator>
p1 Group number
p2 Trip line number (1 to 4)
p3 Trip line display ON/OFF
p4 Display position (0 to 100)
p5 Display color (RED, GREEN, BLUE,

B.VIOLET, BROWN, ORANGE,
Y.GREEN, LIGHTBLUE, VIOLET, GRAY,
LIME, CYAN, DARKBLUE, YELLOW,
LIGHTGRAY, PURPLE, BLACK, PINK,
L.BROWN, L.GREEN, DARKGRAY, OLIVE,
DARKCYAN, S.GREEN)

p6 Line width (1, 2, 3)

Query

SL[ p1[,p2]]?

Example Display trip line 1 in red at the 10% position of

group 1. Set the line width to 1.
SL1,1,ON,10,RED,1

Description The width of the waveform display area along the

amplitude axis is assumed be 100%.

Advertising