LumaSense Technologies INNOVA 1412i User Manual

Page 129

Advertising
background image

Chapter 11

______________________________________________________________________

_____________________________________________________________________________

BE6025-15

1412i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 129 of 202

Acetone (analytically pure)

Spare fine filter-paper DS 0759

Tweezers with Teflon coating

Cotton buds

Clean rubber gloves

To change the filter-paper:

Switch the power to the monitor “Off” by using the AC mains power

switch on its back panel (press O), and then pull out the plug

connecting the monitor to the AC mains power supply.

Put a rubber glove on the hand, which is to hold the parts of the unit.

Unscrew the internal air-filtration unit from its mounting on the air-

inlet, see

Fig.11.1

. To do this, turn the unit anti-clockwise with your

fingers until the unit is free of the slot.

The unit is removed from the monitor. It is shown in

Fig.11.2

.

Use the tweezers to lift off the retaining disc and used filter-paper,

while holding the handle of the unit between your thumb and fingers.

Refer to Fig.11.4. The retaining disc has two tabs which “lock” it to

the filtration unit.

Fig.11.4 Removing the retaining disc and old filter-paper

Holding the disc with the tweezers, moisten a cotton bud with pure

acetone and use it to clean the surfaces of the disc. Place the disc on

a clean, dry surface.

Caution: Make sure that no cotton fibres remain on the mesh of the

retaining disc otherwise they could be sucked up into, and block the

measurement system when the Monitor is operated.

Hold the handle-end of the unit between your thumb and fingers.

Moisten a cotton bud with pure acetone and use it to clean the

surfaces of the unit. Refer to Fig.11.5. Place the unit, with its handle-

end downwards, on a clean, dry surface.