Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6025-15

1412i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 172 of 202

be cut in half and a tubing should be connected between the two cut-

ends of the teflon tubing using the tube fittings supplied for the

purpose (Fig.12.18).

Fig.12.18 Diagram showing how to use Nafion tubing when span

calibrating with one of the following optical filters: UA0983, UA0984

or UA0985

12.10 Calibration of the Water-vapour Filter – Stand-

alone Use

The water-vapour filter has two different functions. Its most

important function is to enable any measured signal to be

compensated for water-vapour interference. Its other function is to

measure the absolute concentration of water-vapour in air samples.

If you only wish to use the water-vapour filter to compensate for

water-vapour interference, then it only has to be zero-point

calibrated. If you wish to use it to measure the absolute

concentration of water vapour in air samples then it has to be zero-

point calibrated and span calibrated.

It is vitally important to accurately zero-point calibrate the water-

vapour filter as this influences the Monitor’s ability to compensate for

interference from water vapour. This demands that the zero-gas you

use during the zero-point calibration of the filter must be perfectly

dry.