3 symbols used in print-outs of measurement data, Section 8.2.3 – LumaSense Technologies INNOVA 1412i User Manual

Page 98

Advertising
background image

Chapter 8

______________________________________________________________________

_____________________________________________________________________________

BE6025-15

1412i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 98 of 202

the text COMMON MARK and GAS MARK are listed below with a

short explanation of what they mean:

Common Marks:

“P”

(1) The Monitor stopped measuring due to a power

failure but has automatically re-started the

monitoring task again after regaining power;

(2) The Monitor has been “reset” either because of

a detected fault or because RESET was pressed.

“E”

The measurement has been marked by the User.

“O”

This symbol indicates that an operational error was

detected during this measurement. The accuracy of

such a measurement cannot always be guaranteed.

“W”

This symbol indicates that a warning was detected

during this measurement. The accuracy of such a

measurement cannot always be guaranteed.

Gas Marks:

“B”

The Monitor was incapable of calculating this gas

concentration.

“F”

After this measurement the filter carousel was

found to be out of alignment. If the carousel was

only slightly out of alignment then the accuracy of

the measurement has not been affected. The ac-

curacy of such a measurement cannot always be

guaranteed.

“A”

The gas concentration measured during this

measurement cycle was found to exceed the user-

defined alarm limit for this gas.

8.2.3

Symbols Used in Print-outs of Measurement Data

Displayed measurements marked with the COMMON MARKS “O”

and “W”, and the GAS MARKS “A” and ”F”, are printed-out with an

“X” symbol. However, any displayed measurements marked with the

COMMON MARKS “P” and/or “E”, and the GAS MARK “B” are not

printed-out with these symbols. These measurements are marked in

a different way:

Common Marks:

The symbol “P” is replaced by the text “1412i Restarted”. This

indicates that the Monitor stopped and then started-up again

Advertising