Chapter 5

______________________________________________________________________

_____________________________________________________________________________

BE6025-15

1412i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 64 of 202

5.3.3

Response of the Monitor after a Partial Reset

The way in which the Monitor responds after a partial reset is

dependent upon the way it was being operated at the time of the

partial reset (see Table 5.1). For example, if the Monitor was busy

taking a measurement when it was “partially reset” it will complete

the interrupted measurement after the partial reset, continue its

monitoring task and make a special “mark” alongside the first

complete measurement cycle it performs after the reset. These

“marks” are described in

section 8.2.2

and

section 8.2.3

.

Table 5.1 Dependence of Monitor’s response, after a partial reset, to

its operating condition at the time of the partial reset.

Monitor last

used while in…

Task being

performed

Task started after a

PARTIAL RESET

Measurement

mode

A gas measurement

Completes the monitoring task

“Display”

Looking at measurement

results on the display

screen

Measurement results are

shown on the display from the

beginning again

Set-up mode

Changing set-up

parameters which control

the operation of the

Monitor

Measurement results are

shown in the display from the

beginning

Memory mode

Handling measurement

results which are stored

in Display Memory and

Background Memory

Completes any interrupted task

and then measurement results

are shown in the display from

the beginning

“Interface”

Obtaining hard-copies

(that is, print-outs of, for

example, measurement

data)

Print-out is stopped and

measurement results are

shown in the display from the

beginning