3 enabling/disabling self tests, Section 5.4.3, Table – LumaSense Technologies INNOVA 1412i User Manual

Page 68

Advertising
background image

Chapter 5

______________________________________________________________________

_____________________________________________________________________________

BE6025-15

1412i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 68 of 202

Table 5.3 The available choices and the default value of the

Monitor’s “beep”

Display Text

Available Choices

Default Value

ENTER SOUND VOLUME 0 0, 1, 2, 3

1

To set the volume:

1. Press SET-UP.

2. Press S3 S1 S1 S2 .

The following screen picture appears:


3. To change the volume, press

¿

.

4. Use the direction push-buttons to enter the correct volume. If you

make a mistake, press S1 to revert back to the original volume.

5. Press

¿

to accept volume.

6. Press SET-UP to exit the set-up function.

5.4.3

Enabling/Disabling Self Tests

If you do not wish the Monitor to perform regular self tests, including

the hardware test on power-up you can disable it.

1. Press SET-UP S3 S1 S1 S3

2. Press S1, the Monitor no longer performs the hardware test as

part of its self-testing procedure during any kind of reset.

3. Press SET-UP to exit the set-up function.

SOUND VOLUME 1

PRESS ENTER TO CHANGE VALUE

REGULAR SELF - TESTS ?

NO YES