Safety Considerations

______________________________________________________________________

_____________________________________________________________________________

BE6025-15

1412i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 5 of 202

Safety Considerations.

Throughout this manual Monitor is used for:

Photoacoustic Gas Monitor – INNOVA 1412i

The Monitor complies with:

·

EN/IEC 61010-1, 2

nd

Edition: Safety requirements for electrical

equipment for measurement, control and laboratory use.

·

Can/CSA-C22.2 No. 61010-1-04 - Safety Requirements for

Electrical Equipment for Measurement, Control, and Laboratory

Use.

·

UL Std. No. 61010A-1 (2

nd

Edition) - Safety Requirements for

Electrical Equipment for Measurement, Control, and Laboratory

Use.

To ensure safe operation and retain the Monitor in safe condition,

note the following:

EXPLOSION HAZARD!

TO AVOID THE POSSIBILITY OF AN EXPLOSION; MONITORING

OF FLAMMABLE GASES IN EXPLOSIVE CONCENTRATIONS

MUST NEVER BE ATTEMPTED.

Never operate the 1412i Photoacoustic Gas Monitor in

potentially explosive environments.

When monitoring potentially flammable or toxic gases it is essential

that:

·

The instrument itself is placed in a well-ventilated area outside the

potentially hazardous zone.

·

A sufficiently long tube is connected to the air-outlet on the back

panel so that the sampled gas is carried away to the open air or to

an extraction and/or filtration unit.