The Auto option regulates the flushing time depending on the

sampling tube length and the cleanliness of the air-filter (as the filter

become dirtier, the flushing time will increase). The Fixed Time

option is divided in to two parts; chamber flushing time and sample

tube flushing time. Once defined, these times are constant regardless

of the sampling tube length and cleanliness of the filters.

Auto Option

a. Press S1, and the following display appears.


b. Press

¿

.

c. Use the direction keys to define the tube length: ▲ and ▼ to

increase and decrease the numbers respectively; t and u to

move the cursor across the fields.

Change the length of the sampling tube shown on the screen to

the length of the tube you attached to the Monitor in

section 4.2

.

Note: the range of acceptable values for the sampling tube

ranges from 00.00m to 99.00m.


d. Press

¿

to select displayed length.

The following display appears:

The Monitor measures gas concentrations in the absolute unit of

mg/m

3

– a unit which is dependent on temperature. Alternatively,

the Monitor can calculate gas concentrations in parts per million

(ppm) by volume – this unit is independent of the temperature of

the gas. If mg/m

3

is chosen as the gas concentration unit it is

therefore necessary to define the temperature at which you wish the

Monitor to calculate gas concentrations. This pre-defined

temperature is termed the NORMALIZATION TEMPERATURE.


e. Press

¿

to change value.

LENGTH OF SAMPLING TUBE 0.00 m

PRESS ENTER TO CHANGE VALUE

NORMALIZATION TEMPERATURE 20.0 º C

PRESS ENTER TO CHANGE VALUE