Monitor. Screw it back into its mounting. To do this, turn the unit

clockwise with your fingers to tighten until it is in position.

Fig.11.7 Refitting the retaining disc

11.1.2 Changing the Filter-paper in the External Air-

filtration Unit

While you do this, we recommend that you wear clean, rubber

gloves.

Tools and equipment required:

Acetone (analytically pure)

Spare fine filter-paper DS 0759

Tweezers with Teflon coating

Cotton buds

Clean rubber gloves

To change the filter-paper:

Switch off the Monitor.

Put a rubber glove on the hand which is to hold the parts of the unit.

Pull the external fine air-filtration unit off the end of the Teflon

®

sampling tube. Hold the stub-end of the external air-filter unit

between your thumb and fingers, and unscrew the coarse air filter

from the end of the unit. Refer to

Fig.11.3

.

Remove the used (old) filter-paper and the retaining disc from the

unit using the tweezers. Refer to

Fig.11.8

.