1 displayed monitoring task data, Section 8.1 – LumaSense Technologies INNOVA 1412i User Manual

Page 91

Advertising
background image

Chapter 8

______________________________________________________________________

_____________________________________________________________________________

BE6025-15

1412i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 91 of 202

8.1 Displayed Monitoring Task Data

When the Monitor is switched on, but not being operated in any

mode, the screen display text identifies the data stored in its Display

Memory, by displaying the day and time of the first stored

measurement cycle of the monitoring task, as well as the day and

time of the last measurement cycle of the monitoring task. For

example:

DATA IN DISPLAY MEMORY RECORDED FROM

2014-01-16 09:50 TO 2014-01-16 10:04

If the Monitor has just been fully reset then the display will just

show the current time and report that the monitor is ready for use.

The Set-up parameters describing a monitoring task and the

measurement data collected during the task are stored as a list in

the Display Memory. You can scroll through the set-up parameters as

well as the data collected during the task by using the direction

push-buttons, q to scroll “vertically” downwards, and p to scroll

“vertically” upwards through the list.

Fig.8.1

illustrates how the data

is organised:

Identification of Measurements:

This screen provides the date and time of the first and last

measurement results, which are stored in Display Memory.

Identification of Each Monitored Gas:

This display identifies the gases monitored during the task. By using

the u push-button, one can find the names of all the gases, which

were monitored during the displayed task.

Statistical Analysis of Measurement Results for Each Gas:

Using the individual measured gas concentration stored in Display

Memory, when Store Measurement History is selected, a statistical

analysis is performed which provides a summary of the stored

measurement results. The MAX (maximum), MIN (minimum) gas

concentrations are displayed, and the µ (Mean or Linear Average)

and σ (Standard Deviation) values are calculated and displayed. A

description of these statistical terms can be found in

section 8.2.1

.

During the performance of a monitoring task the statistical analysis is

updated constantly.

List of Individual Measurement Results:

Each of these displays provides the date and time each measurement

cycle was performed and the concentration of the gas whose name

appears on the screen. In some cases an asterisk (*) will be shown

alongside the gas/vapour concentration. When such an asterisk is

Advertising