LumaSense Technologies INNOVA 1412i User Manual

Page 171

Advertising
background image

Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6025-15

1412i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 171 of 202

2. Let the Monitor continue measuring the span gas A until the

temperature is stable. Look at the print-put of the average and

standard deviation measurements for (1) the water-vapour and

(2) the filter “A”. When these values have stabilised for both

filters, span calibration measurements do not need to continue.


3. Press

¿

.

This causes the following text to appear on the screen:

A : µ : 123.8 μV σ : 617 nV 6 40.4 °C

PRESS ENTER WHEN RESULTS ARE STABLE

These results have been shown on the data-log print-out. If they

have stabilised continue to the next Step.

4. Press

¿

.

The text CALCULATING CALIBRATION FACTORS – PLEASE

WAIT appears on the screen.

The Monitor only uses the last six measurement results when

calculating any calibration factor. The conversion factor (in mg/m3

per volt), which is calculated during a span calibration, is the factor

which allows the Monitor to relate the signal produced by the gas to

the concentration of gas in the cell. The cross-interference

factors, which are calculated during cross-compensation

calibration are those factors which allow the Monitor to compensate

for any interference gas A has on the other installed optical filters.

If any calibration data is doubtful or unacceptable an error message

(marked by an asterisk) will appear on the screen after the

calibration factors have been calculated. Whenever an asterisk is

shown on the screen further information about the condition of the

Monitor can be obtained by pressing the INFO button.

Chapter 14

lists all the error messages connected with doubtful

calibration results and gives an explanation of their significance so

that the user can ascertain what action to take if such messages are

displayed after calibration.

12.9.3 Use of Nafion Tubing during Span Calibration of

UA0983, UA0984 and UA0985.

A Nafion tubing (an optional accessory) is required during the span

calibration of UA0983 e.g carbon dioxide (CO

2

),UA0984 e.g. carbon

monoxide (CO) and UA0985 e.g. dinitrogen oxide (N

2

O). The Teflon

tubing connected to attachment point no. 3 of the “Y”-piece should