Chapter 7

______________________________________________________________________

_____________________________________________________________________________

BE6025-15

1412i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 87 of 202

When you start a new monitoring task the following warning is

displayed WARNING: DISPLAY MEMORY WILL BE DELETED to

remind you to store any data that is presently stored in the Display

Memory. If a monitoring task is stopped, any data already stored in

Display Memory will be deleted when a monitoring task is started.

1.

Press MEASURE.

The following text appears on the screen:

Please note: if you do not use one of the “select” keys within a

short time after the above text appears, the Monitor will

automatically stop operating in Measure mode (the light-emitting

diode in the MEASURE key switches off). If this happens, just press

MEASURE again.

The active monitoring task number is the number of the monitoring

task you wish the Monitor to perform.

2.

If the monitoring task number displayed is correct, then go to

step 5.

If the number on the display is not correct then press S3 to

change the task number.

This changes the display text to the following:

3.

Press S1.

This causes the following text to appear on the display:

ACTIVE TASK : MONITORING TASK NUMBER 1

START TASK DELAYED TASK CHANGE TASK

SELECT TASK

MONITORING TASK CALIBRATION TASK

MONITORING TASK SET-UP NUMBER 1

PRESS ENTER TO CHANGE VALUES