LumaSense Technologies INNOVA 1412i User Manual

Page 16

Advertising
background image

Chapter 2

______________________________________________________________________

_____________________________________________________________________________

BE6025-15

1412i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 16 of 202

5.

Gently lift off the upper cover of the 1412i by pulling it firmly

upward.

6.

Use a screwdriver to remove the fuse-holder:

7.

Remove the old fuse from the holder and replace it with a new.

Check that the fuse has the correct rating.

The fuse must be:

Slow-blow (T) fuse with a rating of 2.5 A, 250 VAC, 5x20 mm, UL

approved.

LumaSense order number: VF0102A

8.

Install the fuse-holder and re-assemble the 1412i again.

The figure below shows the inside of the top cover of the 1412i. Note

that the front side of the cover has square flanges. Place the top