LumaSense Technologies INNOVA 1412i User Manual

Page 80

Advertising
background image

Chapter 7

______________________________________________________________________

_____________________________________________________________________________

BE6025-15

1412i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 80 of 202

The range of acceptable values and the default value of the

averaging period are shown in Table 7.6.

Table 7.6 The range of acceptable values and the default value of the

averaging period

Display Text

Acceptable

Values

Default Value

ENTER AVERAGING PERIOD

00:10

hours:minutes

00:10 to 12:00

hours:minutes

00:10

Stand-alone Use

When the averaging function is active it will not only affect the

results displayed on the screen but also the measurement results

which are being printed out from the Display Memory (see

Chapter

10

). However, when the averaging function is deactivated – by

pressing AVERAGE again (the LED will switch off) – the original

results (un-shaded) will replace the averaged results on the screen.

Gas:

Gas Setup: there are up to six selections here: A, B, C, D, E and W.

These represent the filters installed in the monitor’s filter carousel.

When setting up a new monitoring task, you can choose freely

between the gases available for each filter (names entered when

calibrating the filters, see sections

12.3

and

12.5

). Once a gas has

been selected for a filter and the measurement task started, you can

no longer change the gas for this filter in this measurement task.

However, you can still select gases for filters, which have not

previously been used in the measurement task.

Note: only one gas per filter can be selected at a time for

measuring. The W position always represents the water filter

installed in the monitor.

Sample Integration Time: this enables you to make a choice

between speed and accuracy. Generally, the longer the integration

time the more accurate the results. There are several selection

possibilities here:

Integration Type

Integration Time

Fast

1 s

Normal

5 s

Low Noise

20 s

Advanced

0.5s, 1s, 2s, 5s, 10s, 20s, 50s

Advertising