2 display the homepage of the gas monitor – LumaSense Technologies INNOVA 1412i User Manual

Page 22

Advertising
background image

Chapter 2

______________________________________________________________________

_____________________________________________________________________________

BE6025-15

1412i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 22 of 202

10. Here the IP address of the 1412i, which has been assigned by the

DHCP server, is displayed. This IP address can be used to

manually set Ethernet communication in a PC application, which

accesses the 1412i gas monitor.

Press the S2 (ACCEPT) key to accept.

11. This concludes the setup of the Ethernet parameters of the 1412i

gas monitor.

Press the SET-UP key to leave the setup.

2.5.2

Display the Homepage of the gas monitor

The homepage of the 1412i gas monitor can be displayed if the gas

monitor is connected to an Ethernet network.

In

section 2.5.1

it is explained how to set-up the Ethernet

parameters in the 1412i gas monitor.

Any web browser can be used to display the homepage of the gas

monitor. The IP address obtained from the Ethernet set-up of the

1412i gas monitor can be entered in the address field of the web

browser in order to display the homepage.

The last measured gas values are displayed together with

information about the gas monitor, like the device serial number, if

water- and cross-compensation is performed.

Error and warning messages (see

Chapter 15

for further details) will

also be displayed on the homepage.

The grey "Update Measurements" link can be used to update the

homepage with the last measurement results.

CURRENT DHCP ADDRESS 192.168.000.062