Chapter 7

______________________________________________________________________

_____________________________________________________________________________

BE6025-15

1412i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 84 of 202

Environment branch

13. Press S3 to enter the Environment branch.

14. Use the select push-buttons to select the desired flushing

mode.

If you select Auto, go to step 17.

If you select Fixed Time, go to step 15.

15. Use the direction push-buttons to define the Chamber Flushing

Time.

16. Use the select push-buttons to select Tube Flushing.

If you don’t want tube flushing, select NO and go to step 18.

If you select tube flushing, use the direction push-buttons to

define the flushing time and go to step 18.

17. Use the direction push-buttons to select the desired sample

tube length and move on to the next display.

18. Use the direction push-buttons to select the desired

normalization temperature.

See

section 7.1

for details about the normalization options.

When step 18 is completed, the display returns to the top of the

MEASUREMENT branch.

Press p to go to the top of the Set-up Tree.

If you want measurement data presented as averaged concentration

values, then go to step 20. If you want your measurement data

presented as sample concentration values, then press SET-UP to

exit the set-up function.

Format branch

If you are starting here, press SET-UP:

19. Press S2 to enter the FORMAT branch.

SELECT MEASUREMENT SET-UP BRANCH

MONITORING TASK ENVIRONMENT