1412i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 168 of 202

Fig.12.17 Schematic diagram showing how to set-up a combined

cross-interference and span calibration task (follow the arrows)

If the span gas used during a span calibration task absorbs any of

the light from the other installed optical filters, then it is advisable to

select to perform a cross-interference calibration. This means

that during span calibration the signal in the cell is not only

measured with the water-vapour filter and the filter being span