4 the basic calibration set-up – LumaSense Technologies INNOVA 1412i User Manual

Page 147

Advertising
background image

Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6025-15

1412i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 147 of 202

12.4

The Basic Calibration Set-up

The general equipment required to perform the calibration is shown

in Fig.12.5. Three different lengths of teflon tubing are connected to

a “Y”-piece. The tube attached to point 1 is connected to the air-inlet

of the Monitor; the tube attached to point 2 is attached to a gas flow

meter or similar apparatus; and the tube attached to point 3 is at-

tached to the cylinder of gas which is to be used during calibration.

Fig.12.5 General equipment required for a calibration task

Attaching a gas flow meter serves two vital functions (see Fig.12.6):

Fig.12.6 Schematic diagram of a flow meter