PID TUNING

PID Values determine the performance

and operation of your proportional control

valve. These terms dictate control speed,

control stability, overshoot and oscillation.

All units leave the factory with a generic

tuning designed to handle most applications.

If you encounter issues with valve stability,

oscillation or speed, fine tuning these

parameters may resolve the problem.
Alicat controllers allow you to adjust the

Proportional, Integral and Differential

terms of the PID control loop.
To change the PID loop parameters, push

the button below PID.
Press LOOP TYPE. Then use the LN-UP and

LN-DN buttons to select the appropriate

PID control algorithm. Press SET.
See the following page for descriptions

of the PID Loop Types (PID Control

Algorithms).
P
refers to the Proportional term of the

PID loop.
I refers to the Integral term of the PID loop.
D refers to the Differential term of the PID

loop.
Press P, I or D. Then use SELECT to choose

the digit with the arrow and the UP and

DOWN buttons to change the value. Press

SET to record your value. Press CLEAR to

return to zero.

Before changing the P, I or D

parameter, please record the initial

value so that it can be returned to

the factory setting if necessary.
Valve tuning can be complex. If you

would like assistance, please contact

Alicat for technical support.

I

00000

BACK

LOOP

TYPE

MAIN

P

00100

D

02501

LN-DN

CANCEL

SET

LN-UP

> PD PID

PD2I PID

LOOP

VAR

ON

AUTO

PID

MAIN

SETPT

SOURCE

SETPT

+0.0

Overview of PID Adjustment on Alicat MFCs and Pressure Controllers

http://www.alicat.com/support/instructional-videos/

Advertising