INPUT SIGNALS

Analog Input Signal
Apply analog input to Pin 4 as shown on page 8.
For 6 Pin Locking Industrial Connector, DB9 and DB15 pin-outs see pages 72 to 84.

For PROFIBUS pin-outs see page 70.
Standard 0-5 Vdc is the standard analog input signal. Apply the 0-5 Vdc input

signal to pin 4, with common ground on pin 8. The 5.12 Vdc output on pin 2 can

be wired through a 50K ohm potenti ometer and back to the analog input on pin

4 to create an adjustable 0-5 Vdc input signal source as shown below.

7

6

5

4

3

2

1

8

5.12 Vdc

50 KOhm

Potentiometer

0-5 Vdc

Simple method for providing set-point to controllers

Opti onal 0-10 Vdc: If specifi ed at ti me of order, a 0-10 Vdc input signal can be

applied to pin 4, with common ground on pin 8.
Opti onal 4-20 mA: If specifi ed at ti me of order, a 4-20 mA input signal can be

applied to pin 4, with common ground on pin 8.
NOTE: This is a current sinking device. The receiving circuit is essenti ally a 250

ohm resistor to ground.
NOTE: 4-20mA output requires at least 15 Vdc power input.

CAUTION! D� ��� ������� ���� ������ �� “���� �������’”

�������, �� ���� ���� ������� �������� �� ��� ��������� ��� ����

��� ��������. I� ��� ���� ��������� ���� �������� ���� �������

�������, ������ ��� � ������ �������� ��� � �������� ����� ������.

Electrical Connecti ons and Basic Wiring