Alicat MC Series Mass Flow Controller User Manual

Page 86

Advertising
background image

86

USE of Alicat instruments (M, MW, MS, MC, MCW, MCS, MCR, MCRW, MCRS, P, PS, PC,

PCS, PCR and PCRS product families only) in Class 1 Division 2 applications.

CSA certifies the use of this product for general use as well as use in hazardous

locations as defined by Class 1 Division 2 Group A, B, C and D T4.

CSA certification is indicated by the product label as shown below and

not by the statements in this, or any accompanying documentation.

Special Conditions:

To comply with CSA certification the following information is included in the product literature:

When equipment is properly labeled, it is suitable in Class I, Division 2, Group A, B,

C and D, T4

o Tamb. -40°C to +50°C

Electrical Rating 24Vdc, 0.800A max

Instruments shall be powered by a CSA certified, UL listed, Class II external power

supply suitable for the application

Instruments shall be housed in an enclosure with a minimum IP54 rating or location

providing equivalent protection

Instrument’s final approval shall be provided by the local authority having jurisdiction

EEx nA IIC T4

Class I, Div. 2 Group A, B, C and D T4

24 Vdc, 0.800A max

Class I, Zone 2 AEx nA IIC T4

X – See manual for special conditions

WARNINGS:

EXPLOSION HAZARD – DO NOT

DISCONNECT WHILE CIRCUIT IS LIVE

UNLESS AREA IS KNOWN TO BE NON-

HAZARDOUS.

EXPLOSION HAZARD – SUBSTITUTION

OF COMPONENTS MAY IMPAIR

SUITABILITY FOR CLASS I, DIVISION 2.

www.AlicatScientific.com

Alicat Scientific, Inc.

Tucson, AZ USA

Tel: 520-290-6060

II 3 G

USE of Alicat instruments (M, MW, MS, MC, MCW, MCS, MCR, MCRW, MCRS, P, PS, PC,

PCS, PCR and PCRS product families only) in applications requiring ATEX Certification.

Properly labeled Alicat instruments comply to the following ATEX standard:

II 3 G EEx nA IIC T4 (-40°C ≤ Ta ≤ +50°C)

The examination certificate was issued by the CSA in accordance with

accepted practices and procedures. This confirms compliance with the

European ATEX Directive or Group II Category 3G equipment.

ATEX certification is indicated by the product label as shown above and not by the

statements in this, or any accompanying documentation.
Special Conditions:

Properly labeled equipment is only certified for use in ambient temperatures in the

range of -40°C to +50°C only

Electrical Rating 24Vdc, 0.800A max

Instruments shall be powered by a CSA certified, UL listed, Class II external power

supply suitable for the application

Instruments shall be housed in an enclosure with a minimum IP54 rating or location

providing equivalent protection

Instrument’s final approval shall be provided by the local authority having jurisdiction