Do not connect RS-485 to RS-232 units or cables. Damage will

occur! Check part number or contact factory to verify RS-485
functionality.

Due to variance in cable manufacturing, please identify proper

wiring/pins via continuity check & color when using blunt cut

multi-strand cables.

DB9A Pin-out

The following pin-out chart describes the safest and generally compatible

arrangement when connecting a non-Alicat DB9 wire to a DB9A equipped Alicat.

Not all features may be available between brands, but the common denominators

are featured in our DB9N offerings, along with some options for customization.



7641 N Business Park Dr. Tucson, AZ 85743

520-290-6060 Ph. 520-290-0109 Fax



7641 N Business Park Dr. Tucson, AZ 85743

520-290-6060 Ph. 520-290-0109 Fax

Female Connector Front View

9

9

Male Connector Front View

6

1

6

1

5

5

If your instrument was ordered with a DB9 connection, be sure to check the

Calibration Label on the device and reference the appropriate pin-out diagram.

Pin

Function

1

Not Connected

2

0-5 Vdc Output Signal (or 0-10 Vdc optional)

3

Power In (+Vdc)

4

Ground (common for power, digital communications, analog

signals and alarms)

5

Serial RS-232TX or RS-485 (+)

6

Analog Input Signal [4-20mA, 5Vdc, or 10Vdc] (short to ground

for remote tare function on non-controllers)

7

Ground (common for power, digital communications, analog

signals and alarms)

8

Ground (common for power, digital communications, analog

signals and alarms)

9

Serial RS-232RX or RS-485 (-)

Note: The above pin-out is applicable to all the flow meters and controllers

with the DB9A connector. The availability of different output signals

depends on the options ordered. Optional configurations are noted on the

unit’s calibration sheet.