Alicat MC Series Mass Flow Controller User Manual

Page 41

Advertising
background image

41

Meter does not agree with another meter I have in line.

Volumetric meters are affected by pressure drops. Volumetric flow meters

should not be compared to mass flow meters. Mass flow meters can be

compared against one another provided there are no leaks between the two

meters and they are set to the same standard temperature and pressure. Both

meters must also be calibrated (or set) for the gas being measured. M Series

mass flow meters are normally set to Standard Temperature and Pressure

conditions of 25 ° C and 14.696 psia. Note: it is possible to special order meters

with a customer specified set of standard conditions. The calibration sheet

provided with each meter lists its standard conditions.

When performing this comparison it is best to use the smallest transition

possible between the two devices. Using small transitions will minimize lag and

dead volume.
RS-232 / RS-485 Serial Communications is not responding.

Check that your meter is powered and connected properly. Be sure that the

port on the computer to which the meter is connected is active. Confirm that

the port settings are correct per the RS-232 instructions in this manual (Check

the RS-232 / RS-485 communications select screen for current meter readings).

Close Hyperterminal® and reopen it. Reboot your PC. See pages 10, 11 and 30

for more information on RS-232 / RS-485 signals and communications.
Slower response than specified.

MC-Series Controllers feature a programmable Geometric Running Average

(GRA). Depending on the full scale range of the meter, it may have the GRA

set to enhance the stability/readability of the display, which would result in

slower perceived response time. Please see “Pressure Averaging” and “Flow

Averaging” on page 24.
Jumps to zero at low flow.

MC-Series Controllers feature a programmable zero deadband. The factory

setting is usually 0.5% of full scale. This can be adjusted between NONE and

3.2% of full scale. See page 24.
Discrepancies between old and new units.

Please see “Standard Gas Data Tables” explanation on page 37.

Advertising