RS-232 / RS-485 Output and Input

Configuring HyperTerminal®:
1. Open your HyperTerminal® RS-232 / RS-485 terminal program (installed under

the “Accessories” menu on all Microsoft Windows® operating systems).

2. Select “Properties” from the file menu.

3. Click on the “Configure” button under the “Connect To” tab. Be sure the

program is set for: 19,200 baud (or matches the baud rate selected in the

RS-232 / RS-485 communications menu on the meter) and an 8-N-1-None (8

Data Bits, No Parity, 1 Stop Bit, and no Flow Control) protocol.

4. Under the “Settings” tab, make sure the Terminal Emulation is set to ANSI

or Auto Detect.

5. Click on the “ASCII Setup” button and be sure the “Send Line Ends with Line

Feeds” box is not checked and the “Echo Typed Characters Locally” box and

the “Append Line Feeds to Incoming Lines” boxes are checked. Those settings

not mentioned here are normally okay in the default position.

6. Save the settings, close HyperTerminal® and reopen it.
Streaming Mode (RS-485 units do not have a streaming mode)
In the default Polling Mode, the screen should be blank except the blinking cursor. In

order to get the data streaming to the screen, hit the “Enter” key several times to clear

any extraneous information. Type “*@=@” followed by “Enter” (or using the RS-232 /

RS-485 communication select menu, select @ as identifier and exit the screen). If data

still does not appear, check all the connections and COM port assignments.
Streaming Mode – Advanced
The streaming data rate is controlled by register 91. The recommended default rate

of data provision is once every 50 milliseconds and this is suitable for most purposes.
If a slower or faster streaming data rate is desired, register 91 can be changed to a

value from 1 millisecond to 65535 milliseconds, or slightly over once every minute.
Below approximately 40 milliseconds, data provision will be dependent upon

how many parameters are selected. Fewer data parameters can be streamed

more quickly than more. It is left to the user to balance streaming speed with

number of parameters streamed.
To read register 91, type “*r91” followed by “Enter”.
To modify register 91, type “*w91=X”, where X is a positive integer from 1 to

65535, followed by “Enter”.
To return to the recommended factory default streaming speed, type “*w91= 50”.
Changing From Streaming to Polling Mode:
When the meter is in the Streaming Mode (RS-485 units do not have a streaming

mode), the screen is updated approximately 10-60 times per second (depending

on the amount of data on each line) so that the user sees the data essentially in

real time. It is sometimes desirable, and necessary when using more than one

unit on a single RS-232 line, to be able to poll the unit.
In Polling Mode the unit measures the flow normally, but only sends a line of data

when it is “polled”. Each unit can be given its own unique identifier or address.