MISCELLANEOUS

Miscellaneous is accessed by pressing the MISC button on the Select Menu display.
Next select either MISC1 or MISC2.

MISC1 will display as shown at left.
ZERO BAND refers to Display Zero Deadband.

Zero deadband is a value below which the display

jumps to zero. This deadband is often desired to

prevent electrical noise from showing up on the

display as minor flows or pressures that do not

exist. Display Zero Deadband does not affect the

analog or digital signal outputs.
ZERO BAND can be adjusted between 0 and

3.2% of the sensor’s Full Scale (FS).
Press ZERO BAND. Then use SELECT to choose

the digit with the arrow and the UP and DOWN

buttons to change the value. Press SET to record

your value. Press CLEAR to return to zero.
Pressure Averaging and Flow Averaging may

be useful to make it easier to read and interpret

rapidly fluctuating pressures and flows. Pressure

and flow averaging can be adjusted between 1

(no averaging) and 256 (maximum averaging).
These are geometric running averages where

the number between 1 and 256 can be

considered roughly equivalent to the response

time constant in milliseconds.

This can be effective at “smoothing” high frequency process oscillations such as

those caused by diaphragm pumps.
Press PRESS AVG. Then use SELECT to choose the digit with the arrow and the UP and

DOWN buttons to change the value. Press SET to record your value. Press CLEAR to

return to zero.

Press FLOW AVG. Then use SELECT to choose the

digit with the arrow and the UP and DOWN buttons

to change the value. Press SET to record your value.

Press CLEAR to return to zero.
Setting a higher number will equal a smoother

display.
LCD CONTRAST: The display contrast can be

adjusted between 0 and 30, with zero being the

lightest and 30 being the darkest. Use the UP

and DOWN buttons to adjust the contrast. Press

SET when you are satisfied. Press CANCEL to

return to the MISC display.

PRESS

AVG

BACK

LCD

CONTRAST

MAIN

ZERO

BAND

FLOW

AVG

DOWN

CANCEL

CLEAR

SET

UP

SELECT

0.0

>

DOWN

CANCEL

RESET

SET

UP