Alicat MC Series Mass Flow Controller User Manual

Page 45

Advertising
background image

45

Opti on: Remote Electronics for High Line or Gas Temperatures

Some applicati ons involve operati ng

temperatures outside the standard Alicat

device specifi cati ons. A soluti on using

remote electronics is available. (This opti on is

not applicable for liquid devices.)
The fl ow body’s components are minimized

to only the required sensors. The fl ow data is

sent to the microprocessor electronics up to

6 feet away from the sensor package.
Relocati ng the sensiti ve electronics allows

for installati on of the fl ow body in ambient

temperatures as high as 85° Celsius with gas temperatures under 100°Celsius.
In these applicati ons we recommend our custom gauge calibrati on at a

gas temperature of up to 70°Celsius. This will reduce zero shift errors that

occur when actual gas fl ow temperatures deviate substanti ally from the gas

calibrati on temperature.
This confi gurati on is also used in integrati ons that require a compact fl ow

package at the installati on point.

Opti on: Remote Panel Display

Our Remote Display opti on off ers the fl exibility of using

Alicat’s display with units that are embedded inside

processes or instrument enclosures.
The Remote Display retains all of the same features as our

standard display.
The Remote Display is ideal for:

● OEMs Remote Panel Mounti ng ● Embedded Systems
● Gas Panels

● Fuel Cell Test Stati ons

● Leak Detecti on Systems

● Arti fi cial Environments