Db15h – pin-out “hastings h” style, Db15 pin-outs – Alicat MC Series Mass Flow Controller User Manual

Page 83

Advertising
background image

83

DB15H – Pin-Out “Hastings H” Style

Pin Number

Function

1

N/C

2

RS-232 RX (receive) or RS-485 –

*

3

N/C

4

N/C

5

Ground

6

Primary Analog Signal Output

7

Power Supply Common

8

N/C

9

N/C

10

Secondary Analog Signal Output / fi xed 5.12Vdc

*

11

Power Supply (+Vdc)

12

Ground

13

N/C

14

Analog Tare (meters — when grounded)

Analog Set-Point Input (controllers)

15

RS-232 TX (send) or RS-485 +

*

Check your device’s calibration certifi cate and user manual for the actual electrical input/

output requirements, as all instruments are custom confi gured to some extent.
NOTE: Pins 5, 7 and 12 are connected together inside of the device and are common

grounding points.
N/C = Not Connected/Open (can be used for custom pin assignments – please consult factory).
* Added to allow for full use of features on Alicat devices, may not be present on host wiring.

Male Connector Front View

Female Connector Front View

11

15

14

10

2

6

11

15

14

10

2

6

7

7

The following pin-out chart describes the safest and generally compatible arrangement when

connecting a non-Alicat DB15 wire to a

DB15H equipped Alicat. Not all features may be

available between brands, but the common denominators are featured in our DB15 offerings,

along with some options for customization.

DB15 Pin-Outs

If your instrument was ordered with a DB15 connection,

be sure to check the Calibration Label on the device and

reference the appropriate pin-out diagram.

Advertising