Air Temperature and Relative Humidity Probe
HMP155 Maintenance

Visual Check

Every three months, check that the temperature and humidity probe
mounting is secured and that the cable is in good condition.

Cleaning

Clean the probe with a soft, lint-free cloth moistened with mild detergent.

Changing the Probe Filter

Every six months, change the probe filter:

NOTE

Depending on the conditions of the measurement site, such as air
pollution and dust, the probe filter may need to be changed more
frequently.

NOTE

When handling the probe, do not touch the sensor heads (numbers 3 and
4 in Figure 109 on page 181).

1.

Remove the probe from inside the radiation shield.

1.

Carefully remove the filter from the probe: screw it counter-
clockwise, the pull it out without touching the sensor heads.

2.

After removing the filter, check the O-ring and change it if
necessary. See Figure 109 on page 181. Do not touch the sensor
heads.

3.

Without delay, install the new filter carefully on the probe without
touching the sensor heads.

New filters can be ordered from Vaisala (order code: 219452SP, includes
sintered Teflon filter and O-ring).