Chapter 6 ______________________________________________________________ Maintenance

VAISALA ______________________________________________________________________ 183

Pyranometer CMP6

Periodic Maintenance

The pyranometer is an all-weather instrument. Once installed, the
pyranometer needs little maintenance. It is recommended that you clean
the glass dome of the sensor as part of a regular routine, using water or
alcohol. Check the fixture and leveling of the sensor as needed.

A periodic check is recommended to ensure that the pyranometer is level
and that the silica gel desiccant in the drying cartridge is still colored
orange. When the silica gel is turned from orange to completely
transparent (normally after several months), it must be replaced by fresh
silica gel.

When changing the desiccant in the drying cartridge (indicated with an
arrow in Figure 110 below):

- Make sure the surfaces of the pyranometer and the drying cartridge

that touch the rubber O-ring are clean. Corrosion, dirt, and water can
damage the sensor.

- The rubber O-ring is coated with a silicon grease to improve the seal.

If the rubber O-ring looks dry, apply grease or Vaseline to it.

- Check that the drying cartridge is tightly threaded into the sensor

body.

1012-214

Figure 110 Pyranometer CMP6, Arrow Points to Drying

Cartridge

O-ring type: 24 × 3.0 NBR (nitrite rubber).