Aalborg TIO Totalizer I/O Flow Monitor/Controller User Manual

Page 32

Advertising
background image

28

NOTE: Before enabling the Totalizer, ensure that all Totalizer settings
are configured properly. Totalizer Start values must be entered in
the currently active Volumetric or Mass flow engineering unit. The
Totalizer will not totalize until the Process Flow Rate becomes equal
to or more than the Totalizer Start value. Totalizer Event values must
be entered in currently active volume or mass based engineering
units. If the Totalizer Event at preset total volume feature is not
required, set Totalizer Event value to zero (default settings).

,

b) Totalizer #1 Flow Start (Numerical entry)

This option allows the start of the Totalizer at a preset flow rate. Totalizer
#1 will not totalize until the process flow rate becomes equal to or more
than the Totalizer #1 Flow Start value. The limit of required Totalizer #1
Flow Start value can be entered in increments of 0.1% from 0 - 100%F.S.

c) Totalizer #1 Action Volume (Numerical entry)

This option allows the user to activate preset required action when the
Totalizer reaches a preset volume. Totalizer #1 Action Volume value must
be entered in currently active volume / mass-based engineering units.
Totalizer #1 Action Event becomes true when Totalizer #1 reading is more
or equal to preset "Totalizer #1 Action Volume”. If the Totalizer#1 Action
at preset total volume feature is not required, set “Totalizer #1 Action
Volume” value to zero (default settings).

d) Totalizer #1 Power On Delay (Numerical entry)

Sometimes it is convenient to start the Totalizer only after specified
power-up delay interval. Most of the mass flow meters and controllers
require some warm-up time from the power-up event in order to stabilize
the process variable output and get an accurate reading. “Totalizer #1
Power On Delay” option allows set specified a time interval which must
elapse from the device power-up event before the Totalizer will be
activated. Valid settings are in the range of 0 to 3600 seconds (default
value is 0, no delay).

e) Totalizer #1 Auto Reset (Tabular entry)

This option allows automatic reset of the Totalizer #1 when it reaches
preset Action Volume value. This feature may be convenient for batch
processing when predefined volume of the fluid must be repeatedly
dispensed into the process. The following selections are available:
Enabled or Disabled.