TABLE 6.2 J2 INPUT / OUTPUT JUMPER CONFIGURATION OPTIONS

FOR GFM SERIES FLOW METERS

PV INPUT

TYPE

(TIO INPUT)

J2 JUMPER CONFIGURATION

GFM

CABLE KIT

NOTE

J2A

J2B

J2C

J2D

J2E

J2F

0 - 5 Vdc

2 - 3

5 - 6

8 - 9 10-11 14-15 17-18

KIT-TM-DD

KIT-TM-FD

5 - 10 Vdc

2 - 3

5 - 6

8 - 9 11-12 14-15 17-18 KIT-TM-RD

+5Vdc

reference

signal must

be used

0 - 10 Vdc

2 - 3

5 - 6

8 - 9 11-12 14-15 17-18

NOT

SUPPORTED

FOR GFM

Special Order

option! (PCB

hardware must

be changed)

4 – 20 mA

2 - 3

5 - 6

8 - 9 10-11 13-14 16-17

N/A (user

custom cable

assembly)

(249 Ohm

passive, not

isolated current

input)

d) Parameters Configuration

Following parameters must be configured:

Device Function (See 4.3.12 Submenu “Device Function”). “Meter”

function must be selected.

Full Scale Range (See 4.3.14 Submenu “Device Calibration”). Full

Scale Range parameter must be set equal to the GFM full scale flow
rate in Litr/min.

Fluid Std. Density (See 4.3.14 Submenu “Device Calibration”). This

parameter is required only when mass-based engineering units are
selected.

User may configure other parameters (see Paragraph 5.3) according
to individual preferences and application requirements.

NOTE: If “Full Scale Range”, “Device Function” and “Fluid Std.
Density” parameters are not set properly the device may have
erroneous reading and unpredictable behavior.

,