Aalborg TIO Totalizer I/O Flow Monitor/Controller User Manual

Page 90

Advertising
background image

87

COMMAND

NAME

DESCRIPTION

NO.

COMMAND SYNT

A

X

COMMAND

ARGUMENT 1

ARGUMENT 2

ARGUMENT 3

ARGUMENT 4

RESPONSE

Signal

Conditioner

Settings

Sets/Reads Signal Conditioner

Parameters

Argument1 = F:

Display/Change NR Filter

Parameters:

NRF Sample Number [1-32]

NRF T

ime Limit [0-199]

NRF Error Limit [0.0-10.0%]

16

SC

M

Read/Change Device

Signal Conditioner

Mode

N – No Conditioning

F – NRF Filter

A – Running A

verage

<New Mode>

[N, F

, A]

SCM:<value>

Example:

SCM:F

No Argument

(Returns Current

Mode)

SCM:<value>

Example:

SCM:F

D

Flow Running

A

verage Damping

[0-500 ms]

0 - Disabled

<new value>

in ms [0-500]

SCD:<value>

Example: SCD:50

No Argument

(Returns Current set.)

PWD:<value>

Example: SCD:50

F

NR Filter Settings:

NRF Sample Numb.

NRF T

ime Limit

NRF Error Limit

<new value>

Sample#

[1-32]

<new val.>

T

ime Limit

[0-199]

<new val.>

Error Limit

[0.0-10%]

Example:

PWF:4,8,0.0

No Argument

(Returns Current

set.)

Example:

PWF:4,8,0.0

L

Device Flow

Linearizer

E- Enable

D- Disable

<New V

alue>

E or D

SCL:<value>

Example: SCL:E

No Argument

(Ret. Current

settings)

SCL:<value>

Example: SCL:E