Aalborg TIO Totalizer I/O Flow Monitor/Controller User Manual
Page 89
86
COMMAND
NAME
DESCRIPTION
NO.
COMMAND SYNT
A
X
COMMAND
ARGUMENT 1
ARGUMENT 2
ARGUMENT 3
ARGUMENT 4
RESPONSE
K-Factor
Read and set K-Factor relative to
N2
Mode:
D
–
Disabled (K-Factor = 1.0)
I
–
Internal
U –
User
-Defined
14
K
S
(read current settings)
Returns current mode,
index and UD value
KS:Mode,Index,
UDvalue
Example:
KS:U,1,0.91200
D
(Set Mode: Disabled)
KD
I
(Set Mode: Internal Index)
<Index>
[1-22]
KI:Index,GasName
Example: KI:1,Ar
U
(Set Mode: User
-Defined V
alue)
<V
alue>
[0.00001-999.9]
KU:V
alue
Example: KU:0.91200
Calibration
Settings
Sets/Reads Calibration related
variables.
Argument 1 = T
Hours since last time unit was
calibrated.
NOTE: must be reset to zero after
calibration.
15
C
P
Flow Power Up Delay
[seconds]
<V
alue>
[seconds]
CP:<value>
Example: CP:3
No Argument (Returns
Current Power Up Delay)
CP:<value>
Example: CP:3
F
Device Full Scale Range in
liter/min
<V
alue>
[liter/min]
CF:<value>
Example CF:1000.0
No Argument (Returns
Current Meter FS Range)
CF:<value>
Example: CF:1000.0
L
Device Low Flow Cut Off
in % of full scale
<V
alue>
(%FS)
CL:<value>
Example: CL:5.0
No Argument
(Returns Current value)
CL:<value>
Example: CL:5.0
T
Read/Reset Calibration/
Maintenance T
imer
No Argument
(read timer)
CT
:<value>
Example: CT
:1024.2
Z
Set Cal. T
imer to Zero
CT
:Z