Aalborg TIO Totalizer I/O Flow Monitor/Controller User Manual

Page 83

Advertising
background image

80

COMMAND

NAME

DESCRIPTION

NO.

COMMAND SYNT

A

X

COMMAND

ARGUMENT 1

ARGUMENT 2

ARGUMENT 3

ARGUMENT 4

RESPONSE

Flow Alarms

Sets / reads the parameters of

the Flow Alarms.

Note: If device function is set

to “Meter” the High Alarm

value must be more than Low

Alarm value.

Meter Alarm conditions:

Flow

High Limit

= H

Flow

Low Limit

= L

Low < Flow < High

= N

Controller Alarm conditions:

Flow-SP

High Limit = H

SP-Flow

Low Limit

= L

Low > SP-Flow < High

= N

8

A

C

(set Alarm

configuration)

<V

alue>

(high limit, %F

.S.)

<V

alue>

(low limit, %F

.S.)

AC:40.1,20.5

A

(action delay in

sec.)

<V

alue>

[0-3600]

AA:<V

alue (sec)>

E

(enable Alarm)

A:E

D

(disable Alarm)*

A:D

R

(read

current status)

AR:N (no Alarm)

AR:H (High Alarm)

AR:L (Low Alarm)

S

(Read

current settings)

AS:M,H,L,D,B

where: M – mode (E/D)

H – High settings value

L – Low settings value

D – Action Delay (sec)

B – Latch mode (0-1)

Example:

AS:E,40.0,20.0,2,0

L

(Latch mode)

<V

alue>

(0-disabled*)

(1-enabl’d)

AL:<V

alue> where:

V

alue = 0 – 1

Example:

AL:0