Aalborg TIO Totalizer I/O Flow Monitor/Controller User Manual

Page 85

Advertising
background image

82

COMMAND

NAME

DESCRIPTION

NO.

COMMAND SYNT

A

X

COMMAND

ARGUMENT 1

ARGUMENT 2

ARGUMENT 3

ARGUMENT 4

RESPONSE

To

talizers

Sets and controls action of

the flow T

o

talizers.

NOTE: Start T

o

talizer at

Flow value must be entered

for in %FS (0.0 – 100.0).

Limit volume must be

entered in currently selected

EU. If T

o

talizer hit limit

event is not required, set

“Limit V

o

lume” value (argu-

ment 4) to zero. T

o

talizer#1

can be configured to Count

Up only (argument M is not

supported). T

o

talizer#2 can

be configured to Count Up

or Count Down mode.

When configured to Count

Down mode, limit volume

must be more than zero.

To

talizer#1 reading is stored

in EEPROM (non-volatile)

memor

y. Power cycle will

not affect T

o

talizer#1 read-

ing.T

otalizer#2 reading is

backed up in EEPROM with

6 minutes inter

val. Power

cycle may affect T

o

talizer#2

reading. T

o

talizer#1 cannot

be reset if ResetLock (EE

index 70) value set to 1.

11

T

1 (T

otalizer #1)

2 (T

otalizer #2)

Z

(Reset to zero)

T1Z or T2Z

C

Start flow and Event

Condition

<value>

(start T

o

talizer at flow)

%FS [0.0 – 100.0]

<value>

(Limit volume In current

volume based EU)

T1C:2.5, 0.0

or T2C:2.0,20580.5

P

Power On Delay

<value>

(0-3600 sec.)

T1P:10 or T2P:20

D (disable T

otalizer)*

T1:D or T2:D

E (enable T

o

talizer)

T1:E or T2:E

R (read current T

o

talizer

volume)

T1R:<value> or

T2R:<value> (in current EU)

S

(read current

settings status)

T1S:Mode,Direction,Start

Flow

,LimitV

olume,

PowOnDelay

,AutoResetMode,

AutoResetDelay Example:

T1S:E,0,0.5, 2045.2,10,0,5

A

Set Auto Reset/Reload

mode

<value> [0-1]

0 – Disable

1 – Enable

T1A:0 - disabled

Or T2A:1 - enabled

I

Set Auto Reset/Reload

Inter

val delay

<value>

[0-3600 sec.]

T1I:2 Or T2I:0

B

Restore T

o

talizer#1

value from EE backup

T1B

Not supported by T

o

talizer #2.

M

To

talizer#2 counting

direction configuration

<value> [0-1]

0- Count Up

1-Count Down

T1M:0

Not supported by T

o

talizer #1.